Omnigrad M TC12
Modular TC thermometer

Omnigrad M TC12
Modular TC thermometer ©Endress+Hauser

Comprehensive and common temperature measurement technology for almost all branches of industry

The high modular and robust thermometer finds its use in several applications in almost all industries. An optional head transmitter, with all common communication protocols, makes the device ready to use with enhanced measurement accuracy and reliability compared to directly wired sensors. A variety of process connections, dimensions and materials offer flexible application possibilities.

  • Avantajlar

    • High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion lengths

    • High degree of insert compatibility and design as per DIN 43772

    • Fast response time with reduced/tapered tip form

    • ...

  • Uygulama alanı

    • Universal range of application

    • Measuring range: -40 to 1100 °C (-40 to 2012 °F)

    • Pressure range up to 40 bar (580 psi)

    • Degree of protection: up to IP68

Özellikler ve şartlar

Geri

Details

  • Measuring principle

    Thermocouple

  • Characteristic / Application

    metric style

    modular temperature assembly

    universal range of application

    suitable for hazardous areas

    process connection as compression fitting

    without neck

    incl. thermowell / protection tube (metal)

  • Terminal head

    TA20A - aluminium - cover with screws

    TA20B - plastics - screw cover

    TA20D - aluminium - hinged cover high

    TA20J - SS - deep-drawn w. or w/o. display

    TA20R - SS - machined w. screw cover

    TA20W - aluminium - cover w. snap fit

    TA21E - aluminium - screw cover w. chain

    TA30A - E+H, aluminium - cover low - w. or w/o. display

    TA30D - E+H, aluminium - cover high

  • Thermowell / protection tube

    welded protection tube

  • Insert / probe

    mineral insulated (MI), flexible

  • Outer diameter protection tube / Insert

    9,0 mm (0,35'')

    11,0 mm (0,43'')

    12,0 mm (0,47'')

  • Max. immersion length on request

    up to 10.000,00 mm (393,70'')

  • Material protection tube/ thermowell

    1.4404 (316L)

    1.4571 (316Ti)

    Alloy 600 (2.4816)

    Alloy C276 (2.4819)

  • Optional coating

    Not defined

  • Process connection

    compression fitting:

    G1/2"

    G1"

    R1/2"

    R3/4"

  • Tip shape

    straight

    reduced

    tapered

  • Surface roughness Ra

    0,8 μm (31,5 μin.)

  • Operating temperature range

    Type K:

    -40 °C ...1.100 °C

    (-40 °F ...2.012 °F)

    Type J:

    -40 °C ...750 °C

    (-40 °F ...1.382 °F)

  • Max. process pressure (static)

    at 20 °C: 40 bar (580 psi)

  • Accuracy

    class 1 acc. to IEC 60584

  • Response time

    depending on configuration

    t50 = 6 s

    t90 = 13 s

  • Integration head transmitter

    yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

  • Ex - approvals

    ATEX II

    ATEX IECEx

    NEPSI

    IECEx
    EAC Ex

  • Certification

    Gost Metrology

    SIL (transmitter only)

Dokümanlar / Kılavuzlar / Yazılım

Aksesuarlar / Yedek parçalar

Nr
Order code
Specification