Aramayı başlatmak için en az 2 karakter girin.

Digital conductivity sensor
Condumax CLS15D

Digital conductivity sensor
Condumax CLS15D ©Endress+Hauser
Digital conductivity sensor
Condumax CLS15D ©Endress+Hauser
Digital conductivity sensor
Condumax CLS15D ©Endress+Hauser

The conductivity of a liquid can be measured using the conductive or toroidal measuring principles. This video shows what it is about and how these measuring principles work.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens conductive conductivity sensor for standard applications in pure and ultrapure water

Condumax CLS15D is the digital solution for conductivity measurement in the low measuring ranges. It performs reliably and accurately in a wide range of applications - even in hazardous areas. Designed for low maintenance and a long operating life, the sensor offers best value for money. With Memosens digital technology, CLS15D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration, and facilitates predictive maintenance.

  • Basit ürünler

  • Seçmesi, kurması ve çalıştırması kolay

Teknik mükemmellik

Sadelik

  • Standart ürünler

  • Güvenilir, sağlam ve az bakım gerektiren

Teknik mükemmellik

Sadelik

  • Üst düzey ürünler

  • Son derece fonksiyonel ve uygun

Teknik mükemmellik

Sadelik

  • Özelleştirilmiş ürünler

  • Zorlu uygulamalar için tasarlanmış

Teknik mükemmellik

Sadelik

Değişken

FLEX seçimleri Teknik mükemmellik Sadelik
  • F
  • L
  • E
  • X

Fundamental seçim

Temel ölçüm ihtiyaçlarınızı karşılayın

Teknik mükemmellik
Sadelik
  • F
  • L
  • E
  • X

Lean seçim

Ana proseslerinizin üstesinden kolayca gelin

Teknik mükemmellik
Sadelik
  • F
  • L
  • E
  • X

Extended seçim

Proseslerinizi yenilikçi teknolojilerle optimize edin

Teknik mükemmellik
Sadelik
  • F
  • L
  • E
  • X

Xpert seçim

En zorlu uygulamalarınızda uzmanlaşın

Teknik mükemmellik
Sadelik

Değişken

  • Özelliklere genel bakış

    • Measurement range

      k=0,01: 0.04 to 20 µS/cm
      k=0,1: 0.10 to 200 µS/cm

    • Process temperature

      Threaded with fixed cable:
      -20 to 100 °C (-4 to 212 °F)
      Threaded with plug-in head:
      -20 to 120 °C (-4 to 248 °F)
      Sterilization: max. 140 °C (284 °F) for 30 minutes

    • Process pressure

      13 bar at 20 °C (188 psi at 68 °F)
      1 bar at 120 °C (14 psi at 248 °F)

please wait Loading alternative products

Hızlı seçim

Önerdiğimiz seçeneklerden birini seçin
Satış fiyatı
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

veya

Ayrı konfigürasyon

Seçiminize mümkün olan tüm seçeneklerle devam edin
  • Avantajlar

    • Reliable and accurate measuring values at low conductivities

    • Easy to clean thanks to polished measuring surfaces

    • Sterilizable and autoclavable

    • Quality certificate stating the individual cell constant

    • Logging of sensor-specific data for easy traceability and...

  • Uygulama alanı

    Condumax CLS15D measures conductivity in pure and ultrapure water applications for the monitoring and control of:

    • Ion exchangers

    • Reverse osmosis

    • Cooling water

    • Distillation

    • Chip cleaning

    The digital sensor can be used in power plants for:

    • Degassed...

Özellikler ve şartlar

Geri

Details

  • Measuring principle

    Conductive

  • Application

    Measurement in pure and ultrapure water range, monitoring of ion exchangers, reverse osmosis, distillation, chip cleaning

  • Characteristic

    Analog or digital 2-electrode conductivity sensors

  • Measurement range

    k=0,01: 0.04 to 20 µS/cm
    k=0,1: 0.10 to 200 µS/cm

  • Measuring principle

    Conductive conductivity cell with polished stainless steel electrodes

  • Design

    2-electrode conductivity cell with coaxially arranged electrodes, electropolished

  • Material

    Electrode: stainless steel 1.4435
    Electrode shaft: PES

  • Dimension

    Electrode diameter: 16 mm (0.63 inch)
    Electrode length: appr. 55 mm (2.17 inch)

  • Process temperature

    Threaded with fixed cable:
    -20 to 100 °C (-4 to 212 °F)
    Threaded with plug-in head:
    -20 to 120 °C (-4 to 248 °F)
    Sterilization: max. 140 °C (284 °F) for 30 minutes

  • Process pressure

    13 bar at 20 °C (188 psi at 68 °F)
    1 bar at 120 °C (14 psi at 248 °F)

  • Temperature sensor

    NTC 30K

  • Ex certification

    ATEX, NEPSI, FM, CSA

  • Connection

    Process: 1/2" and 3/4" NPT, Clamp 1 1/2"
    Cable: Memosens connector

  • Ingres protection

    IP68

  • Additional certifications

    Calibration certification of the cell constante, material certification 3.1

Dokümanlar / Kılavuzlar / Yazılım

Aksesuarlar / Yedek parçalar

Nr
Order code
Specification
    • Order code 50004381

      PVDF Adapter G 1/2" - G 1"

      for mounting of sensors with thread NPT 1/2".

    • Order code 50004738

      AM 32 equalizing sleeve DN32, PVC

      for conductivity cell installation into T-90-piece or cross piece DN 32

    • Order code 50004739

      AM 40 equalizing sleeve DN40, PVC

      For conductivity cell installation into T-90-piece or cross piece DN40

    • Order code 50004740

      AM 50 equalizing sleeve DN50, PVC

      For conductivity cell installation into T-90-piece or cross piece DN50

    • Order code 50081902

      Calibration solution CLY11-A, 74,0 uS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO

    • Order code 50081903

      Calibration solution CLY11-B,149,6 uS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO.

    • Order code 50081904

      Calibration solution CLY11-C,1,406mS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO

    • Order code 50081905

      Calibration solution CLY11-D,12,64 mS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM to NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO

    • Order code 50081906

      Calibration solution CLY11-E,107,00mS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO

    • Order code CYA21-

      Flowfit CYA21

      Stainless steel flow-through assembly for mounting on analytical panels. Construction of analytical panels for applications within powerplants and the (petro-) chemical industry among others. Fits for applications up to 100oC at 16 bar. Robust stainless steel assembly; fit for applications at higher pressure and temperature rates; perfect for bypass installations; inclusive mounting kit.

    • Order code 50066536

      PVC-Threaded coupling G 1/2"

      for cementing in cross / T-pieces DN 20