TP analyzer
Liquiline System CA80TP

Liquiline System CA80TP - TP analyzer for environmental monitoring ©Endress+Hauser
Inside view of the TP analyzer with cooling module ©Endress+Hauser

The total phosphorus concentration of a liquid can be determined using chemical and thermal digestion and the colorimetric measuring principle. This video shows how this measuring principle works.

Total phosphorus analyzer for environmental monitoring, industrial and municipal wastewater

Phosphorus is the decisive factor for excessive algae and plant growth in water bodies, that's why continuous monitoring of this parameter is essential. The Liquiline System TP analyzer offers precise online measurement of total phosphorus and ensures highest process safety. Automatic calibration and cleaning save you operating costs while advanced diagnostics with remote access support you in providing process documentation to the authorities.

  • Avantajlar

    • Molybdenum blue method according to EN ISO 6878:2004 guarantees direct comparability to laboratory results

    • Meeting industry needs: The analyzer is perfectly suited for environmental monitoring, industrial and municipal wastewater

    • Fast and easy process...

  • Uygulama alanı

    The Liquiline System TP analyzer monitors:

    • Inlet and outlet of wastewater treatment plants for documentation purposes and calculation of the cleaning capacity

    • Discharges of industrial wastewater treatment plants to determine discharge fees and to...

Özellikler ve şartlar

Geri

Details

  • Measuring principle

    Colorimetric

  • Characteristic

    Analyzer for total phosphorus

  • Size

    Cabinet:
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 inch
    Stand housing:
    1693 x 530 x 417 mm
    66.65 x 20.87 x 16.42 inch

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 to 40 °C (39.2 to 104 °F)

  • Ambient temperature

    5 to 40 °C (41 to 104 °F)

  • Process pressure

    At atmospheric pressure, < 0.2 bar

  • Sample flow rate

    Min. 5 ml/min

  • Consistency of the sample

    Low solid content (TS<50 mg/l), homogeneous

  • Specials

    Easy upgrade to measuring station with up to four digital Memosens sensors
    Automatic calibration and cleaning
    User-configurable measuring, cleaning and calibration intervals

    Self-priming analyzer with optical dosing unit l
    Dilution module (optional)
    Digital communication for remote access

  • Application

    Monitoring and optimization of the cleaning capacity of municipal and industrial wastewater treatment plants
    Monitoring of the wastewater treatment plant outlet for documentation purposes
    Monitoring and optimization of activated sludge basins

  • Power supply

    100 to 120 VAC / 200 to 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 or 60 ± 1,2 Hz

  • Output

    2x 0/4 to 20 mA
    Webserver, Modbus, Ethernet/IP, Profibus DP

  • Input

    1 or 2 measuring channel
    1 to 4 digitale sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    0.05 to 10 mg/l
    0.5 to 50 mg/l

  • Consumables

    Reagents and standard solutions CY80TP are necessary for the operation

Dokümanlar / Kılavuzlar / Yazılım

Aksesuarlar / Yedek parçalar

Nr
Order code
Specification