Aramayı başlatmak için en az 2 karakter girin.

Digital ORP sensor
Ceraliquid CPS42D

Ceraliquid CPS42D - Memosens ORP glass sensor with refillable liquid KCl electrolyte ©Endress+Hauser
Ceraliquid CPS42D for harsh chemical applications ©Endress+Hauser

The pH-value of a liquid can be calculated using the potentiometric measurement principle. This video shows what it is about and how this measuring principle works.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens glass electrode for applications with fast-changing medium compositions or low conductivity

Ceraliquid CPS42D is the digital high performer for harsh chemical applications, media with low conductivity or a considerable content of organic solvents. The sensor is designed for fast response guaranteeing high process safety even in applications with fast-changing media. Thanks to Memosens digital technology, CPS42D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration and facilitates predictive maintenance.

Hızlı seçim

Önerdiğimiz seçeneklerden birini seçin
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

veya

Ayrı konfigürasyon

Seçiminize mümkün olan tüm seçeneklerle devam edin
  • Avantajlar

    • Resistant to poisoning due to constant refilling of KCl bridge electrolyte and separate reference lead

    • Applicable at very low conductivities (= 5 µS/cm) thanks to liquid KCl electrolyte

    • Suitable for cleaning in place (CIP) and sterilization in place (SIP)

    • Perfectly suited for quickly changing media: Combination of liquid...

  • Uygulama alanı

    Media with very low conductivities or a considerable amount of organic solvents or alcohols and fast changing media:

    • Chemical industry

    • Power plants, e.g. boiler feed water

    • Life sciences, e.g. water for injection

    With ATEX, FM and CSA approval for use in...

Özellikler ve şartlar

Geri

Details

  • Measuring principle

    Sensor ORP / Redox

  • Application

    - Special applications with high requirements with regard to accuracy, speed.- rapidly changing composition of media, highly clogging media, low conductivities.

  • Characteristic

    - Digital electrode with Memosens technology - gel-free, refillable electrolyte - greatest accuracy - can be subject to pressure to prevent clogging

  • Measurement range

    -1500 mV - +1500 mV

  • Measuring principle

    - Liquid filled compact electrode with ceramic diaphragm - platinum ring

  • Dimension

    Diameter: 12 mm
    Shaft lengths: 120, 225 mm

  • Process temperature

    -15°C - 130°C

  • Process pressure

    max. 8bar

  • Ex certification

    ATEX
    FM
    CSA

  • Connection

    Inductive transfer of digital signals.

  • Ingres protection

    IP68

  • Additional certifications

    EHEDG

Dokümanlar / Kılavuzlar / Yazılım

Aksesuarlar / Yedek parçalar

Nr
Order code
Specification